產品簡介
| 產地類別 | 進口 | 應用領域 | 醫療衛生,生物產業 |
細胞-材料相互作用力測試分析儀,細胞材料黏附脫落測試評估系統,細胞與材料間粘合力測試分析系統,細胞黏附材料測試儀,細胞與材料間黏附力檢測分析系統。

1、培養液浸沒式粘附黏合力測試分析
2、支持無損傷倒置探針和倒置探針(ASTM D2979)或環形(ASTM D6195)粘性測試
These tack testing fixtures can be used to measure the tack properties (instantaneous adhesion) of adhesives either through the inverted probe (ASTM D2979) or the loop (ASTM D6195) tack testing methods. It comprises: for the inverted probe technique (a flat indenter with a 5-mm diameter and annular loading weights) and for the loop (a tensile grip and a 1 inch flat base). The force required to completely peel out the material provides its tackiness.
Cell attachment or, lack of it, to biomaterials is a critical factor affecting the performance of a device or implant. Cell attachment is a complicated, time-dependent, process involving significant morphological changes of the cell and deposition of a bed of extracellular matrix. Details of the adhesive bond that is formed have been reviewed by, for example, Pierres et al (2002) (4), Lukas and Dvorak (2004) (5), and Garcia and Gallant (2003) (6). The strength of this coupling can be determined either by monitoring the force of attachment between a cell and a substrate over time or by measuring the force required to detach the cell once it has adhered.
Cell adhesion to a surface depends on a range of biological and physical factors that include the culture history, the age of the cell, the cell type, and both the chemistry and morphology of the underlying surface and time. These elements that need to be considered in developing a test protocol.
Devising robust methods for measuring the propensity of cells to attach to different substrates is further complicated since either cell adhesion or detachment can be assessed. These processes that are not always similar or complementary.
Most studies of cell attachment focus on obtaining some measure of the time-dependent force required to detach, or de-adhere, cells that have already adhered to a surface (James et al, 2005) (7). More recently investigators have begun to measure the adhesive forces that develop between cells and the underlying surface during attachment (Lukas and Dvorak, 2004) (5). From a practical point of view, it is much easier to measure the force required to detach or de-adhere cells from a surface than to measure those that develop during attachment. However, in both cases, the experimental data should be interpreted with a degree of caution that depends on the intended use of the measurements. The methods of measuring cell adhesion described herein are measures of the force required to detach an adherent cell.
The purpose of this guide is to provide an overview of current generic test methods and identify the key factors that influence the assessment of cell adhesion and detachment. It is anticipated that this guide will form the basis for producing a series of standards that will describe these test methods in more detail.
Scope
1.1 This guide describes protocols that can be used to measure the strength of the adhesive bond that develops between a cell and a surface as well as the force required to detach cells that have adhered to a substrate. Controlling the interactions of mammalian cells with surfaces is fundamental to the development of safe and effective medical products. This guide does not cover methods for characterizing surfaces. The information generated by these methods can be used to obtain quantitative measures of the susceptibility of surfaces to cell attachment as well as measures of the adhesion of cells to a surface. This guide also highlights the importance of cell culture history and influences of cell type.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
關聯品牌
Biomomentum是一家專注于組織和生物材料力學性能測試儀的設計和制造的公司。代表性產品Mach-1多軸向組織和材料的力學性測試分析系統Arthro-BST關節軟骨壓電流動電位測試分析系統已經成為組織材料力學和電位測試分析的行業重要儀器。自1999年以來,MACH-1*的多軸力學測試系統已經幫助了世界各地的數百名科學家增進和發布與生物材料、組織和軟材料相關的創新性研究。Biomomentum公司亦提供第三方生物力學測試技術服務。
查看更多留言咨詢
移動站:細胞-材料相互作用力測試分析儀
您訪問的細胞-材料相互作用力測試分析儀產品信息,由世聯博研(北京)科技有限公司自行發布提供,產品價格為面議,如您想了解更多關于細胞-材料相互作用力測試分析儀型號、參數、用途等信息,歡迎咨詢。
該內容的真實性、準確性和合法性由發布會員負責,為保障安全,建議您在購買相關產品前務必確認供應商資質以及產品質量!
相關分類
該廠商其他產品
*您想獲取產品的資料:
個人信息:
用戶評論
發布評論